INFRATHERM pyrometer ISR 50-LO

52

Operation mode:

AAkaX

X = 1 MONO; X = 2 RATIO

Emissivity slope:

AAevXXXX

XXXX = (0800 ... 1200, decimal, equivalent to 0.800 ... 1.200)

Emissivity:

AAemXXXX

XXXX = (0010 ... 1000‰) (decimal)

Exposure time t

90

:

AAezX

X = 0 ... 6 (decimal)

0 = intrinsic time constant of the device

1 = 0.01 s

3 = 0.25 s

5 = 3.00 s

2 = 0.05 s

4 = 1.00 s

6 = 10.00 s

Clear time maximum
value storage:

AAlzX

X = 0 ... 8 (dec.)

0 = Maximum value storage off

1 = 0.01 s

4 = 1.00 s

7 = external deletion

2 = 0.05 s

5 = 5.00 s

8 = automatically deletion

3 = 0.25 s

6 = 25.00 s

9 = hold

External clearing:

AAlx

Simulation of an external deletion contact

Analog output:

AAasX

X = 0...1

0 = 0...20 mA 1 = 4...20 mA

Reading basic tem-
perature range:

AAmb

Output: XXXXYYYY (hex 8-digit, °C)

XXXX = beginning of temperature range
YYYY = end of temperature range

Reading tempera-
ture sub range:

AAme

Output: XXXXYYYY (hex 8-digit, °C)

XXXX = beginning of temperature range
YYYY = end of temperature range

Setting of tempera-
ture sub range:

AAm1XXXXYYYY XXXX (hex 4-digit) beginning of temp. range (°C)

YYYY (hex 4-digit) end of temp. range (°C)

Reading switch-of
limit:

AAaw or AAar

Output: XX (decimal) 02 … 50 in %

Switch-of limit:

AAawXX

XX (decimal) 02 … 50 in %

Reading warning
level dirty window:

AAdw

Output: XX (decimal) 00 … 99 in %

Setting warning
level dirty window:

AAdwXX

XX (dec.) ** … 99 in %, 00 … < ** switches off this function
( ** is the set value for the switch-off limit „aw“)

Address:

AAgaXX

XX = (00 ... 97)

00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting com-
mands!)

Baud rate:

AAbrX

X = 0 ... 6 or 8 (dec.)

0 = 1200 baud

3 = 9600 baud

6 = 57600 baud

1 = 2400 baud

4 = 19200 baud

(7 is not allowed)

2 = 4800 baud

5 = 38400 baud

8 = 115200 baud

Changing °C / °F

AAfhX

Output: X = 0: display in °C; X = 1: display in °F

Wait time:

AAtwXX

XX = 00 ... 99 (decimal)

Internal
temperature:

AAgt

Output: XX (dec. 00 ... 98, in °C)

XXX (dec. 032 ... 208°F)

Max. internal
temperature:

AAtm

Output: XX (dec. 00 ... 98, in °C)

XXX (dec. 032 ... 208°F)

Reading signal
strength

AAtr

Output: XXXX, XXXX = 0000 ... 1500 in ‰
(calculated signal strength of the measuring object)

Error status:

AAfs

Output 1 byte hex (00 = no error)

Bit 0 = 1: Measurement unit doesn’t work
Bit 1 = 1: Internal temperature measurement doesn’t work

Laser targeting light: AAlaX

X = 0 switch off laser; X = 1 switch on laser

Reading interface:

AAin

Output: 1 or 2 (1 = RS232, 2 = RS485)

Lock keyboard:

AAlkX

X = 0 ... 3

1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3

Reading parame-
ters:

AApa

Output decimal 11-digit:

Digit 1 and 2 (10...99 or 00): Emissivity

ε

Digit 3 (0 ... 6): Exposure time t

90

Digit 4 (0 ... 8): Clear time max. storage t

clear

Digit 5 (0 ... 1): Analog output
Digit 6 and 7: (00 ... 98): Internal temperature (°C)
Digit 8 and 9 (00 ... 97): Address