15 reference numbers, 1 reference numbers instruments, 2 reference numbers accessories – LumaSense Technologies ISR 50-LO Benutzerhandbuch

Seite 53: Reference numbers, Reference numbers instruments, Reference numbers accessories

Advertising
background image

INFRATHERM pyrometer ISR 50-LO

53

Digit 10 (0 ... 6 or 8): Baud rate
Digit 11 (0): Always 0
Digit 12 … 15: Emissivity slope

Device type:

AAna

Output: “ISR 50-LO“ (16 ASCII-characters)

Serial number:

AAsn

Output: XXXX (hex 4-digit)

Ref. number:

AAbn

Output: XXXXXX (hex 6-digit)

Device type /
software version:

AAve

Output: XXYYZZ (6-digit decimal)

XX = 54 for ISR 50-LO
YY = Month of software version
ZZ = Year of software version

Software version
in detail:

AAvs

tt.mm.yy XX.YY

tt = day; mm = month; yy = year; XX.YY = software version

Note: the letter “l” means the lower case letter of “L”.

Additional instruction for the RS 485 interface:
Requirements to the master system during half-duplex operation:

1. After an inquiry, the bus should be switched into a transmission time of 3 bits

(some older interfaces are not fast enough for this).

2. The pyrometer's response will follow after 3 ms at latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated.

15

Reference numbers


15.1

Reference numbers instruments

3 882 900

ISR 50-LO MB 18:

700 ... 1800°C

3 882 910

ISR 50-LO MB 25:

800 ... 2500°C

3 882 920

ISR 50-LO MB 30:

1000 ... 3000°C


Scope of delivery: Converter ISR 50-LO with HD fibre, length: 2,5 m (other length 5 m, 6 m, 10 m or 15 m
possible for extra charge) and optical head, works certificate, PC software „InfraWin“, user manual.

Ordering note:
When ordering the following data are necessary:
• the measuring distance the optical head has to be adjusted

• the length of the fibre in case of another length as the standard of 2.5 m

A connection cable is not included in scope of delivery.


15.2

Reference numbers accessories

3 821 440

Connection cable 5 m, 11 wires, with additional digital cable (1 m)

3 821 450

Connection cable 5 m, 4 wires (for power supply and analog output)

3 834 390

ball and socket mounting for optical head

3 834 230

adjustable mounting support for optical head

3 835 180

air purge for optical head

3 835 240

90° mirror for optical head

3 852 290

Power supply NG DC for DIN rail mounting; 100 … 240 V AC

⇒ 24 V DC, 1 A

3 852 540

power supply NG 0D for DIN rail mounting; 85 ... 265 V AC

⇒ 24 V DC, 600 mA

3 852 550

power supply NG 2D, as NG 0D: additionally with 2 limit switches

3 890 640

LED digital display DA 4000-N

3 890 650

LED digital display DA 4000: with 2 limit switches

3 890 560

LED digital display DA 6000-N: with possibility for pyrometer parameter settings for digital
INFRATHERM pyrometers; RS232 interface

3 890 570

LED digital display DA 6000-N with RS485 interface

3 890 520

LED digital display DA 6000; DA 6000-N additional with 2 limit switches and analog input
and output

3 890 530

LED digital display DA 6000 with RS485 interface

3 826 500

HT 6000, portable battery driven indicator and instrument for pyrometer parameter setting

Advertising