2 measuring circuit monitoring, 3 analog output, Measuring circuit monitoring – JUMO 14597 safetyM TB/TW08 Temperature Limiter and Temperature Monitor Operating Manual Benutzerhandbuch
Seite 100: Analog output, 8 technical data, 2 measuring circuit monitoring 8.3 analog output

8 Technical Data
38
8.2
Measuring circuit monitoring
8.3
Analog output
RTD temperature probe
and KTY11-6
Double thermo-
couples
Thermo-
couples
Current 0 ... 20 mA, 4 ... 20mA
Voltage 0 ... 10 V, 2 ... 10 V
Overrange
and underrange
is detected
LED K1 and KV are lit; "9999“ flashes in the display
Probe and
wire break
is detected
LED K1 and KV are lit; "9999“ flashes in the display; relay output
K1 is inactive
is detected at 4…20mA
and 2…10V
LED K1 and KV are lit;
"9999“ flashes in the display
Relay output K1 is inactive
Probe short-circuit
is detected
LED K1 and KV are lit; "9999“ flashes in the display;
relay output K1 is inactive
is not de-
tected
Type of sig-
nal
Accuracy
Residual ripple
Load influence Temperature coeffi-
cient
Load resistance
Current
4 ... 20 mA
≤
0.5 %
± 0.5 %
± 0.01 mA
80 ppm/K
≤
500
Ω
0 ... 20 mA
Voltage
2 ... 10 V
≤
0.5 %
± 0.5 %
± 15 mV
80 ppm/K
÷
500
Ω
0 ... 10 V