Determining the fitting k factor using a teach-in procedure depending on a volume ("teAch V.")

The device will use the new K factor as soon as "SAVE YES" is confirmed when leaving the Parameters
menu.

Prepare a tank with a known volume.

Stop the fluid circulation.

Confirm "TEACH V.": "FILL END" is displayed.

K-FACtOR

K=2.8500

tEACh V.

REtURN

0......9

The device calculates the K factor of the fitting
and displays it.

FiLL END

0000.0 L

VALiD N/Y

tEACh F.

OUtPUt

Charge the pipe to fill the tank.

When the tank is full, confirm "FILL END":

K=2.9000

Enter the volume (value between 0.1 and 9999.9 in the units
chosen for the totalizers at the "UNIT" parameter) of fluid
that passed in the circuit.

Confirm.

The display shows the K factor of the fitting, may it have been entered
or determined by a Teach-In procedure.

Confirm the displayed value.

Confirm the K factor determined through Teach-In or not.

Figure 31:

Teach-In procedure depending on a volume

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.