Threshold values, Figure 38 – Burkert Type 8025 Benutzerhandbuch

Seite 49

Advertising
background image

47

Operatingandcommissioning

8.6.10 configuring the transistor output Do1 to switch a load

depending on two threshold values

OUtPUt

DO1

hYStERES.

DO2

1-= 0.000

1+= 0.000

Enter a flow rate value

1)

, associated to the high

threshold, in the unit chosen in the "UNIT" parameter.

Confirm.

wiNDOw

Enter a flow rate value

1)

, associated to the low

threshold, in the unit chosen in the "UNIT" parameter.

Confirm.

iNV YES

iNV NO

0......9

Choose the operating, inverted or not inverted, of the
transistor output.

DEL. 1= 00

Enter the delay before switching (value between 0
and 99 s).

Confirm.

REtURN

On a version WITH relay outputs

(see chap. 8.6.12 et 8.6.13.

On a version WITHOUT
relay outputs

FiLtER

1)

If hysteresis operating: value set for 1- ≤ value set for 1+.

If window operating: value set for 1- < value set for 1+.

Figure 38:

Configuration of the DO1 transistor output with switching thresholds

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

English

Type 8025 UNIVERSAL