8.6.16 parameterizing the cut-off flow rate

This parameter makes it possible to set the flow rate value to 0 if the measured value is less than the set cut-off
value:

• the display then shows a flow rate = 0 (a dot is displayed after the flow rate units).

• the outputs and the totalizers react as if the actual flow rate was = 0.

CUt-OFF

CO=0.000

BACKLit

Enter the measured flow rate value below which the device considers that the flow rate
is = 0.

To disable the function, set CO=0.

Confirm.

Figure 47:

Diagram of the "CUT-OFF" parameter of the Parameters menu

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

8.6.17 Setting the brightness of the display and how long it stays

oN, or deactivating the backlight

This parameter makes it possible:

• to adjust the brightness of the display and how long the display is backlit after a key press.

• to deactivate the backlight.

BACKLit

BKLG= 1

END

Set the display brightness, from 1 to 9.

Confirm.

BKLG= 9

BKLG OFF

0......9

Or, deactivate the backlight.

Confirm.

tiME= 30

Set the duration the display is backlit after a key press (00 s: the
display is constantly backlit).

Confirm.

END

Figure 48:

Diagram of the "BACKLIT" parameter of the Parameters menu

English

Type 8025 UNIVERSAL