Operatingandcommissioning

The following table shows the response times (10% to 90%) for the filters:

Filter

Response time

Filter

Response time

0

0,15 s

5

6 s

1

0,7 s

6

10 s

2

1,4 s

7

19 s

3

2,5 s

8

33 s

4

3,5 s

9

50 s

Filter 0: measured
flow rate

Filter 3

Filter 6

Filter 9

slow

fast

Figure 45:

Available filters

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

8.6.15 Resetting both totalizers

This parameter makes it possible to reset both totalizers.

Both totalizers are reset upon confirmation of "SAVE YES" when leaving the Parameters menu.

tOtAL

RESEt N

CUtOFF

0......9

Choose to reset the two totalizers or not.

Confirm

RESEt Y

Figure 46:

Diagram of the "TOTAL" parameter of the Parameters menu

The daily totalizer can be reset from the Process level. See chap. 8.5.

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.