Proline Promass 80 PROFIBUS PA

Installation

15

a0003597

Fig. 7:

Installation in a vertical pipe (e.g. for batching applications)

1 = Supply tank , 2 = Sensor, 3 = Orifice plate, pipe restrictions (see Table), 4 = Valve, 5 = Batching tank

System pressure

It is important to ensure that cavitation does not occur, because it would influence the oscillation
of the measuring tube. No special measures need to be taken for fluids which have properties similar
to water under normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction
lines, it is important to ensure that pressure does not drop below the vapor pressure and that the
liquid does not start to boil. It is also important to ensure that the gases that occur naturally in many
liquids do not outgas. Such effects can be prevented when system pressure is sufficiently high.

For this reason, the following installation locations are preferred:
• Downstream from pumps (no danger of vacuum)
• At the lowest point in a vertical pipe.

1

2

3

4

5

DN

Ø Orifice plate, pipe

restrictor

DN

Ø Orifice plate, pipe

restrictor

mm

inch

mm

inch

1

1/24"

0.8

0.03

40

1½"

22

0.87

2

1/12"

1.5

0.06

40 FB

1½"

35

1.38

4

1/8"

3.0

0.12

50

2"

28

1.10

8

3/8"

6

0.24

50 FB

2"

54

2.00

15

½"

10

0.40

80

3"

50

2.00

15 FB

½"

15

0.60

100

4"

65

2.60

25

1"

14

0.55

150

6"

90

3.54

25 FB

1"

24

0.95

250

10"

150

5.91

FB = Full bore versions of Promass I