Troubleshooting

Proline Promass 80 PROFIBUS PA

72

• For FAILSAFE_TYPE

→ LAST GOOD (factory setting):

If a valid output value was available before the failure:

If no valid output value was available before the failure:

• For FAILSAFE_TYPE

→ WRONG VALUE:

For status information, see the table in the following section.

!

Note!
The FAILSAFE_TYPE function can be configured in the corresponding Analog Input function block
1 to 6 or Totalizer function block 1 to 3 by means of an operating program (e.g. FieldCare).

9.2.2

List of system error messages

Quality code (HEX)

Quality status

Quality substatus

Limits

0x44
0x45
0x46

UNCERTAIN

Last usable value

OK

Low

High

Quality code (HEX)

Quality status

Quality substatus

Limits

0x4C
0x4D

0x4E

UNCERTAIN

Initial Value

OK

Low

High

No.

Device status message
(local display)

PROFIBUS measured value status

Extended
diagnostic
message in the
PROFIBUS
master

Cause/remedy (spare part

Page 80 ff.)

Q

u

a

li

ty co

de (H

EX

)

M

e

as

ur

ed value s

tat

us

Qu

ali

ty

st

a

tu

s

Q

u

alit

y s

ubs

ta

tu

s

Li

m

its

Depicted on the local display:
S = System error
$ = Fault message (with an effect on the outputs)
! = Notice message (without any effect on the outputs)

001

S: CRITICAL FAILURE
$: # 001

0x0F

BAD

Device Failure

Constant

ROM / RAM
failure

Cause:
ROM/RAM error. Error when accessing the
program memory (ROM) or random access
memory (RAM) of the processor.

Remedy:
Replace the amplifier board.

011

S: AMP HW EEPROM
$: # 011

0x0F

BAD

Device Failure

Constant

Amplifier EEPROM
failure

Cause:
Amplifier with faulty EEPROM

Remedy:
Replace the amplifier board.

012

S: AMP SW EEPROM
$: # 012

0x0F

BAD

Device Failure

Constant

Amplifier EEPROM
data inconsistent

Cause:
Error when accessing data of the measuring
amplifier EEPROM

Remedy:
Perform a "warm start" (= start the measuring
system without disconnecting main power).

Access:
SUPERVISION

→ SYSTEM → OPERATION →

SYSTEM RESET (

→ RESTART)