10 technical data, 1 technical data at a glance, 1 applications – FMC Corporation - Talstar 80 Benutzerhandbuch

Seite 88: 2 function and system design, Measuring principle, Measuring system, 3 input, Measured variable, Measuring range, Measuring ranges for liquids (promass f, m)

Advertising
background image

Technical data

Proline Promass 80 PROFIBUS PA

88

10

Technical data

10.1

Technical data at a glance

10.1.1

Applications

→ Page 7

10.1.2

Function and system design

Measuring principle

Mass flow measurement by the Coriolis principle

Measuring system

→ Page 9

10.1.3

Input

Measured variable

• Mass flow (proportional to the phase difference between two sensors mounted on the measuring

tube to register a phase shift in the oscillation)

• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)

Measuring range

Measuring ranges for liquids (Promass F, M):

Measuring ranges for liquids (Promass E, H, S, P):

DN

Range for full scale values (liquids)

g

min(F)

to

g

max(F)

[mm]

[inch]

8

3/8"

0 to 2000 kg/h

0 to 73.5 lb/min

15

1/2"

0 to 6500 kg/h

0 to 238 lb/min

25

1"

0 to 18000 kg/h

0 to 660 lb/min

40

1 ½"

0 to 45000 kg/h

0 to 1650 lb/min

50

2"

0 to 70000 kg/h

0 to 2570 lb/min

80

3"

0 to 180000 kg/h

0 to 6600 lb/min

100*

4"*

0 to 350000 kg/h

0 to 12860 lb/min

150*

6"*

0 to 800000 kg/h

0 to 29400 lb/min

250*

10"*

0 to 2200000 kg/h

0 to 80860 lb/min

*) only Promass F

DN

Range for full scale values (liquids)

g

min(F)

to

g

max(F)

[mm]

[inch]

8

3/8"

0 to 2000 kg/h

0 to 73.5 lb/min

15

1/2"

0 to 6500 kg/h

0 to 238 lb/min

25

1"

0 to 18000 kg/h

0 to 660 lb/min

40

1 ½"

0 to 45000 kg/h

0 to 1650 lb/min

50

2"

0 to 70000 kg/h

0 to 2570 lb/min

Advertising