Troubleshooting

Proline Promass 80 PROFIBUS PA

74

380

S: UPPER FREQ. LIM
$: # 380

0x0F

BAD

Device Failure

Constant

Frequence limit 2

Cause:
The measuring tube oscillation frequency is outside
the permitted range.

Causes:
Measuring tube damaged
Sensor defective or damaged

Remedy:
Contact your Endress+Hauser service organization.

381

S: FLUIDTEMP.MIN.
$: # 381

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Fluid Temperature
Min.

Cause:
The temperature sensor on the measuring tube is
likely defective.

Remedy:
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is

correctly plugged into the amplifier board

• Remote version: Check sensor and transmitter

terminal connections No. 9 and 10
→ Page 29 ff.

382

S: FLUIDTEMP.MAX.
$: # 382

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Fluid Temperature
Max.

383

S: CARR.TEMP.MIN
$: # 383

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Carrier
Temperature Min.

Cause:
The temperature sensor on the carrier tube is likely
defective.

Remedy:
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is

correctly plugged into the amplifier board

• Remote version: Check sensor and transmitter

terminal connections No. 11 and 12
→ Page 29 ff.

384

S: CARR.TEMP.MAX
$: # 384

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Carrier
Temperature Max.

385

S: INL.SENS.DEF.
$: # 385

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Inlet Sensor Defect

Cause:
One of the measuring tube sensor coils (inlet or
outlet) is likely defective.

Remedy:
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is

correctly plugged into the amplifier board

• Remote version: Check sensor and transmitter

terminal connections No. 4, 5, 6, 7.

386

S: OUTL.SENS.DEF.
$: # 386

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Outlet Sensor
Defect

387

S: SEN.ASY.EXCEED
$: # 387

0x10
0x11
0x12

BAD

Sensor Failure

O.K.

Low

High

Sensor Asymmetry
Exceeded

388

S: AMP. FAULT
$: # 388

0x0F

BAD

Device Failure

Constant

Amplifier failure

Cause:
Amplifier error

Remedy:
Contact your Endress+Hauser service organization.

389

S: AMP. FAULT
$: # 389

0x0F

BAD

Device Failure

Constant

Amplifier failure

390

S: AMP. FAULT
$: # 390

0x0F

BAD

Device Failure

Constant

Amplifier failure

No.

Device status message
(local display)

PROFIBUS measured value status

Extended
diagnostic
message in the
PROFIBUS
master

Cause/remedy (spare part

Page 80 ff.)

Q

u

a

lit

y co