10.1.8

Operating conditions: Environment

Ambient temperature range

Standard: –20 to +60 °C (-4 to +140°F) (sensor, transmitter)
Optional: –40 to +60 °C (-40 to +140°F) (sensor, transmitter)

!

Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.

Storage temperature

–40 to +80 °C (–40 to +175 °F) (preferably +20 °C (+68 °F))

Degree of protection

Standard: IP 67 (NEMA 4X) for transmitter and sensor

Shock resistance

According to IEC 68-2-31

Vibration resistance

Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6

CIP cleaning

Yes

SIP cleaning

Yes

Electromagnetic compatibility
(EMC)

To IEC/EN 61326 and NAMUR Recommendation NE 21

10.1.9

Operating conditions: Process

Medium temperature range

Sensor:

Promass F, A, H, P:

–50 to +200 °C (–58 to +392 °F)

Promass F (high-temperature version):

–50 to +350 °C (–58 to +662 °F)

Promass M, I, S:

–50 to +150 °C (–58 to +302 °F)

Promass E:

–40 to +140 °C (–40 to +284 °F)