WIKA operating instructions, models CTH6300 and CTH6500

2079988.04 07/2014 GB/D

GB

6. Commissioning, operation

6.4.1 Unit switching °C and °F or % r. h., td or g/m³ [Unit]
Unit = Einheit

Measuring unit temperature (°C = Celsius, °F = Fahrenheit)

Measuring unit humidity (% rH = relative humidity, td = dew point, g/m³ = absolute

humidity)

Press the ENTER/MENU key and select Unit using the arrow keys ▲▼. Then press

ENTER/MENU again. A small 1 appears on the left side of the display, it indicates the

channel.

Use the arrow keys ▲▼ to select the channel for which the displayed unit needs to be

changed. Confirm by pressing ENTER/MENU.

Depending on the set probe °C/°F or % rH/td/gm³ is displayed on the right side of the

display (see chapter 6.4.2 “Probe selection Prob”). Select the required unit using the

arrow keys ▲▼ and confirm it by pressing ENTER/MENU.

Press ESC to return to the measuring mode.

6.4.2 Probe selection Prob

Press the ENTER/MENU key and select Prob using the arrow keys ▲▼. Then press

ENTER/MENU again.

A small 1 appears on the left side of the display, it indicates the channel. Using the arrow

keys ▲▼ switch the channel for which a probe should be selected.

Confirm selection by pressing ENTER/MENU. Now select the following probes using

the arrow keys ▲▼ (see table on the next page):

MENU

Unit

Prob

Lin2

CAL

Chnl

ArEA

Lo6

MENU

Unit

Prob

Lin2

CAL

Chnl

ArEA

Lo6