WIKA operating instructions, models CTH6300 and CTH6500
2079988.04 07/2014 GB/D
GB
6. Commissioning, operation
Measurement
parameter
Probe selection (Prob)
LCD display
Temperature
Pt100 (RTD)
Temperature
Fe-CuNi type J
Temperature
NiCr-Ni type K
Temperature
Fe-CuNi type L
Temperature
NiCrSi-NiSi type N
Temperature
Pt13Rh-Pt type R
Temperature
Pt10Rh-Pt type S
Temperature
Cu-CuNi type T
Humidity
% r. h.
Flow
m/s
Pressure
Pa
Hot wire
m/s
Confirm the required setting by pressing ENTER/MENU. Press ESC to return to the
measuring mode again.
Make sure that the correct measurement parameter is set for the
connected probe. If a measurement parameter is changed in the Prob
menu and the change confirmed by pressing the Enter key, the standard
calibration is used automatically.
Note on combined probes (temperature and humidity):
Set relative humidity as a measurement parameter for the measuring
channel to which the combined probe is connected. If you want the
temperature value of the connected combined probe to be displayed as
well, deactivate the channel to which no combined probe is connected
(see chapter 6.4.6 “Activate/deactivate measuring channels [Chnl]”.