6. Commissioning, operation

Use the arrow keys ▲▼ now to choose between rH for humidity calibration and P °C for

temperature calibration.

Use the arrow keys ▲▼ to select the required calibration function.

Standard characteristic curve oFF

Select oFF by means of the arrow keys ▲▼. Press ENTER/MENU to confirm the

required setting.

Press ESC to return to the measuring mode.

Calibration by code oP1

Select oP1 by means of the arrow keys ▲▼. Press ENTER/MENU to confirm the

required setting. A small 1 appears in the lower part of the display. Four characters

follow it (hex code / 0 ... F). Change the 4 characters by means of the arrow key ▲. Press

the arrow key ▼ to go to the next position. If all 4 characters are entered as required,

confirm by pressing ENTER/MENU. A small 2 appears and the following 4 characters

can be changed as well.

Press ESC to return to the measuring mode.