Gb 6. commissioning, operation – WIKA CTH6300 Benutzerhandbuch
Seite 26

26
WIKA operating instructions, models CTH6300 and CTH6500
2079988.04 07/2014 GB/D
GB
6. Commissioning, operation
Example of the measured value display after entering a probe calibration code:
The small 1 in the left corner in connection with the display segment CAL in the center
of the display indicates that the calibration by code oP1 has been activated.
Physical calibration oP2
Use the arrow keys ▲▼ to select oP2. Press ENTER/MENU to confirm the required
setting. 1 P is displayed in the lower display part.
Use the arrow keys ▲▼ to choose from 1-point 1 P-, 2-point 2 P- and 3-point 3 P-
calibration.
Example of 1-point calibration:
Confirm 1-point calibration 1 P by pressing ENTER/MENU. Go appears on the display.
Once the measured value is stable, confirm by pressing ENTER/MENU.
After approx. 2 seconds P1 appears in the first display line for the measured value 1, dP
appears in the 2nd line for the decimal point.
Use the arrow keys ▲▼ to select the required number of decimal places:
dP. = two decimal places
dP . = one decimal place (decimal point moves one position to the right)
oP2
1 P
2 P
3 P