E1oh, E1eh, Dismounting, return and disposal – WIKA CTH6300 Benutzerhandbuch

Seite 37

Advertising
background image

37

WIKA operating instructions, models CTH6300 and CTH6500

2079988.04 07/2014 GB/D

GB

9. Faults / 10. Dismounting, return and disposal

Indication

Cause

Measures

E1oh

Interruption of the Auto-off

function

Switch on the Auto-off function

again, see chapter 6.7 “AUTO-

OFF function”.

E1eh

Interruption of the Auto-off

function

Switch on the Auto-off function

again, see chapter 6.7 “AUTO-

OFF function”.

E23

EE-prom content destroyed

Send in for repair

E25

EE-prom content destroyed

Send in for repair

E12

Overflow

Remove battery and insert it

again

E31

Temperature of the reference

measuring point is outside of the

measuring range

Use the temperature probe

according to the technical

specifications. See chapter

3 “Specifications”.

CAUTION!

If faults cannot be eliminated by means of the measures listed above, the

hand-held thermometer must be shut down immediately.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter

10.2 “Return”.

10. Dismounting, return and disposal

WARNING!

Residual media at the dismounted hand-held thermometer and/or

temperature probe can result in a risk to persons, the environment and

equipment.

Take sufficient precautionary measures.