6. Commissioning, operation
Press ENTER/MENU to confirm the required setting. Si_ appears on the display. Use
the arrow keys ▲▼ to select the sign:
Si_ = the number to be entered is in the negative range (below 0.00 °C)
Si┘ = the number to be entered is in the positive range
Press ENTER/MENU to confirm the required setting. Fd 0 appears on the display.
Select the range using the arrow keys ▲▼:
Fd 0 = lower than 1,000 °C
Fd 1 = higher than 1,000 °C
Press ENTER/MENU to confirm the required setting. 00.00 appears on the display (or
similar). Enter the temperature of your reference now. Change the digits using the arrow
key ▲. Use the arrow key ▼ to switch to the next digit. After completing the input of the
temperature confirm by pressing ENTER/MENU.
Press [ESC] to return to the measuring mode.
The physical calibration oP2 cannot be cancelled by means of the ESC
key. If required, the calibration can be cancelled by switching off the
measuring instrument.