WIKA operating instructions, models CTH6300 and CTH6500

2079988.04 07/2014 GB/D

GB

6. Commissioning, operation

Press ENTER/MENU to confirm the required setting. Si_ appears on the display. Use

the arrow keys ▲▼ to select the sign:

Si_ = the number to be entered is in the negative range (below 0.00 °C)
Si┘ = the number to be entered is in the positive range

Press ENTER/MENU to confirm the required setting. Fd 0 appears on the display.

Select the range using the arrow keys ▲▼:

Fd 0 = lower than 1,000 °C
Fd 1 = higher than 1,000 °C

Press ENTER/MENU to confirm the required setting. 00.00 appears on the display (or

similar). Enter the temperature of your reference now. Change the digits using the arrow

key ▲. Use the arrow key ▼ to switch to the next digit. After completing the input of the

temperature confirm by pressing ENTER/MENU.

Press [ESC] to return to the measuring mode.

The physical calibration oP2 cannot be cancelled by means of the ESC

key. If required, the calibration can be cancelled by switching off the

measuring instrument.