9. Faults

9. Faults

In case of maloperation or faults the instrument helps the operator by means of the

following error messages.

Indication

Cause

Measures

Wrong probe or no probe

connected

Connect probe or connect

correct probe.

“too low” underrange of the

measuring range

Use the temperature probe

according to the technical

specifications. See chapter

3 “Specifications”.

“too high” overrange of the

measuring range

Use the temperature probe

according to the technical

specifications. See chapter

3 “Specifications”.

Temperature of the reference

measuring point exceeds the

upper limit of the measuring

range

Use the temperature probe

according to the technical

specifications. See chapter

3 “Specifications”.

Temperature of the reference

measuring point is below the

lower limit of the measuring

range

Use the temperature probe

according to the technical

specifications. See chapter

3 “Specifications”.

E15

Battery is completely discharged Insert new batteries

E19

Battery is completely discharged Insert new batteries

E1dh

Battery is completely discharged Insert new batteries

E16

Interruption of the Auto-off

function

Switch on the Auto-off function

again, see chapter 6.7 “AUTO-

OFF function”.