WIKA operating instructions, models CTH6300 and CTH6500

2079988.04 07/2014 GB/D

GB

6. Commissioning, operation

6.4.3 Activating/deactivating display option differential temperature [Lin2]

(only for 2-channel instruments)

Press the ENTER/MENU key and select Lin 2 using the arrow keys ▲▼. Then press

ENTER/MENU again. Now activate or deactivate the display “Differential temperature”

T1-T2 using the arrow keys ▲▼.

If T1-T2 is visible on the LCD, the differential temperature is active. Confirm the required

setting by pressing ENTER/MENU.

Press ESC to return to the measuring mode.

To display differential temperature, both channels must be activated.

6.4.4 Calibration mode CAL

This measuring instrument allows to perform a simple calibration when replacing the

probes in order to compensate the manufacturer-specific tolerances of the probes and

guarantee a consistently high accuracy of the measuring chain.

The instrument has 3 different calibration modes:

[OFF]: Standard characteristic curve

for example, for Pt100 resistance measurements DIN IEC 60751

oP1:

Calibration by code

The 2 x 4-digit code is displayed clearly visible on the handles of our probes,

corresponds to a 2-point calibration

oP2: Physical calibration

Calibration with test variables: 1-point, 2-point or 3-point calibration possible

The current calibration coefficients are given in the enclosed traceable calibration

certificate.

CAL = calibrate

Press the ENTER/MENU key and select CAL using the arrow keys ▲▼, then press

ENTER/MENU again. A small 1 appears on the left side of the display, it indicates the

channel. Use the arrow keys ▲▼ to select the channel (1 or 2) to be calibrated.

MENU

Unit

Prob

Lin2

CAL

Chnl

ArEA

Lo6