6. Commissioning, operation
6.4.7 Storage management [Lo6] (not possible with CTH6300)
Press the ENTER/MENU key and select Lo 6 using the arrow keys ▲▼ (only if a logging
device is ordered). Then press ENTER/MENU again. OFF appears in the lower display
line. Use the arrow keys ▲▼ to start the logger mode by means of ON now. Confirm
selection by pressing ENTER/MENU.
Use the arrow keys ▲▼ to choose between the automatic Auto or manual SPot saving
and confirm by pressing ENTER/MENU. Decide whether to add or create a new log file
by selecting Add and nLo6. When selecting the automatic saving at the end you are
asked to select the measuring interval:
1S
= 1 second
1M
= 1 minute
5S
= 5 seconds
2M
= 2 minutes
10S = 10 seconds
5M
= 5 minutes
20S = 20 seconds
10M = 10 minutes
30S = 30 seconds
20M = 20 minutes
Confirm the input by pressing ENTER/MENU. Use ESC to return to the measuring
mode. Manual saving by means of SPot. Press ESC to save individual measured values
manually.
6.5 Storage query [HOLD-MAX-MIN-AVE]
After pressing the HOLD-MAX-MIN-AVE key for the first time, the current measured
values at the moment of actuation are “frozen” and indicated on the display as Hold
values. Press this key repeatedly to request saved maximum, minimum and average
values in the lower line (small display).
Note for 2-channel instruments:
After displaying the hold values for both channels (large and small
display), the MAX-MIN-AVE values of the first channel are displayed in
the lower line (small display), after that the values of the second channel
are displayed. If only one probe is connected to a 2-channel instrument,
the 2nd channel should be deactivated (see chapter 6.4.5 “Activate/
deactivate measuring channels (only for 2-channel instruments [Chnl]”.