EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 375

SCPI-Befehlsreferenz
OSA
367
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:
BANDwidth[1|2]|BWIDth[1|2]:
RelativeLEVel?
Description
This query returns a value indicating either the
current or the minimum/maximum channel
bandwidth position setting for WDM analysis.
At *RST, this value is set to 3.0 dB for bandwidth1
and 20.0 dB for bandwidth2.
Syntax
:CALCulate[1..n][:WDM]:BANDwidth[1|2]|BWI
Dth[1|2]:RelativeLEVel?[<wsp>MAXimum|MINi
mum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<PowerLevel>