Status:operation:bit<8|9>: condition – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 597

SCPI-Befehlsreferenz
OSA
589
Product-Specific Commands—Description
The <Condition> response represents the
current operation condition of the instrument.
The meaning of the response depends on the
value returned for bit <n>.
bit <8>: When the returned value is 1, the
instrument is performing a measurement (trigger
system INITiated).
bit <9>: When the returned value is 1, the
instrument is performing an offset nulling and/or
a wavelength referencing
(CALibration:ZERO:AUTO?).
Example(s)
STAT? Must return READY
CAL:ZERO:AUTO ONCE
STAT:OPER:BIT9:COND? Keep resending this
query as long as the operation is not complete
(returned value is not 0).
See Also
:CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus?
:STATus:OPERation:BIT<8|9>:
CONDition?