EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch

Seite 457

Advertising
background image

SCPI-Befehlsreferenz

OSA

449

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the <Width> parameter.

The <Width> parameter corresponds to a valid

width in hertz for the noise measurement region.

The

CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh

:FREQuency? MIN and

CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh

:FREQuency? MAX queries can be used to

determine valid width values.

Example(s)

CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ

CALC:WDM:CHAN:SEL "ITU_22"

CALC:WDM:CHAN:NOIS:TYPE POLY5

CALC:WDM:CHAN:NOIS:WIDTh:FREQ 75.0 GHZ

CALC:WDM:CHAN:NOIS:WIDTh:FREQ? Returns

7.500000E+010

Notes

Custom width for noise measurement region is

applied only if selected noise type is

POLYnomial5.

See Also

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT

h[:WAVelength]

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT

h:FREQuency?

:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE

:CALCulate[1..n][:WDM]:CHANnel:SELect

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

WIDTh:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:WIDTh:FREQuency

Advertising