SCPI-Befehlsreferenz

OSA

553

Product-Specific Commands—Description

:SENSe[1..n]:AVERage:COUNt

Description

This command sets the number of acquired

traces to combine for averaging to a specific

value.

At *RST, averaging count is set to 8.

Syntax

:SENSe[1..n]:AVERage:COUNt<wsp><Count>|

MAXimum|MINimum|DEFault

Parameter(s)

Count:
The program data syntax for <Count> is defined

as a <numeric_value> element. The <Count>

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the <Count> parameter.

The <Count> parameter corresponds to a valid

averaging count value.

The SENSe[1..n]:AVERage:COUNt? MIN and

SENSe[1..n]:AVERage:COUNt? MAX queries can

be used to determine valid count range.