SCPI-Befehlsreferenz

522

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:OSNR:

BANDwidth|BWIDth[:RESolution]:AUTO

Description

This command controls activation of WDM

analysis OSNR calculation using auto resolution

bandwidth for all channels.

At *RST, auto resolution bandwidth is set to off

(disabled).

Syntax

:CALCulate[1..n][:WDM]:OSNR:BANDwidth|BWI

Dth[:RESolution]:AUTO<wsp><Auto>

Parameter(s)

Auto:
The program data syntax for <Auto> is defined

as a <Boolean Program Data> element. The

<Auto> special forms ON and OFF are accepted

on input for increased readability. ON

corresponds to 1 and OFF corresponds to 0.