EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch

Seite 461

Advertising
background image

SCPI-Befehlsreferenz

OSA

453

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the <Width> parameter.

The <Width> parameter corresponds to a valid

width in meter for the noise measurement

region.

The

CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh

[:WAVelength]? MIN and

CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh

[:WAVelength]? MAX queries can be used to

determine valid width values.

Example(s)

CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ

CALC:WDM:CHAN:SEL "ITU_22"

CALC:WDM:CHAN:NOIS:TYPE POLY5

CALC:WDM:CHAN:NOIS:WIDTh:WAV 12.5 NM

CALC:WDM:CHAN:NOIS:WIDTh:WAV? Returns

1.250000E-008

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:WIDTh[:WAVelength]