SCPI-Befehlsreferenz

OSA

503

Product-Specific Commands—Description

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

The <Select> parameter corresponds to a valid

channel result index to select. The channel index

cannot be zero.

The

CALCulate[1..n][:WDM]:DATA:CHANnel:COUNt?

query can be used to determine valid index

range.

Example(s)

CALC:WDM:CHAN:DEL:ALL

CALC:WDM:CHAN:DEF "C_007", 1380.000 NM

<Do measurement>

CALC:WDM:DATA:CHAN:NSEL 1

See Also

:CALCulate[1..n][:WDM]:CHANnel:NSELect

:CALCulate[1..n][:WDM]:DATA:CHANnel:COUNt?

:CALCulate[1..n][:WDM]:DATA:CHANnel:NSELec

t?

:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect

:CALCulate[1..n][:WDM]:DATA:CHANnel:

NSELect