EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 390

SCPI-Befehlsreferenz
382
FTB-5240S/S-P/BP
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Distance> parameter.
The <Distance> parameter corresponds to a
valid distance in meter from peak to center of
noise region.
The
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance[:WAVelength]? MIN and
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance[:WAVelength]? MAX queries can be
used to determine valid distance values.
Example(s)
CALC:WDM:CHAN:AUTO:NOIS:TYPE POLY5
CALC:WDM:CHAN:AUTO:NOIS:DIST:WAV 40.0
NM
CALC:WDM:CHAN:AUTO:NOIS:DIST:WAV?
Returns 4.000000E-008
Notes
Custom noise measurement distance is applied
only if selected noise type is POLYnomial5.
See Also
:CALCulate[1..n][:WDM]:CHANnel:AUTO
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance[:WAVelength]?
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan
ce[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:DISTance:[WAVelength]