Status – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 595

SCPI-Befehlsreferenz
OSA
587
Product-Specific Commands—Description
:STATus?
Description
This query returns a value indicating the global
status of the instrument.
This command has no associated *RST
condition.
Syntax
:STATus?
Parameter(s)
None
Response Syntax
<Status>
Response(s)
Status:
The response data syntax for <Status> is defined
as a <CHARACTER RESPONSE DATA> element.
The <Status> response represents the
instrument state, where:
UNINITIALIZED means the instrument has not
been initialized yet.
INITINPROGRESS means the instruments
initialization is in progress.
READY means the instrument is ready.
BUSY means the instrument is busy.
DISCONNECTED means the instrument is
disconnected.
DEFECTIVE means the instrument is defective.
Example(s)
STAT?
See Also
:CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus:OPERation:BIT<8|9>:CONDition?