EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch

Seite 393

Advertising
background image

SCPI-Befehlsreferenz

OSA

385

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:AUTO:

NOISe:WIDTh:FREQuency

Description

This command sets the frequency width of the

noise measurement region of the WDM analysis

default channel.

At *RST, the width of the default channel noise

measurement region is set to 100.0 GHz.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

WIDTh<wsp><Width[<wsp>HZ]>|MAXimum

|MINimum|DEFault

Parameter(s)

Width:
The program data syntax for <Width> is defined

as a <numeric_value> element followed by an

optional <SUFFIX PROGRAM DATA> element.

The allowed <SUFFIX PROGRAM DATA>

element is HZ. The <Width> special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

Advertising