Calculate[1..n][:wdm]:channel: noise:auto – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 444

SCPI-Befehlsreferenz
436
FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:AUTO
Description
This command controls activation of i-InBand
noise measurement for WDM analysis of the
selected channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to off (disabled).
Syntax
:CALCulate[1..n][:WDM]:CHANnel:NOISe:AUTO
<wsp><Auto>
Parameter(s)
Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of auto noise measurement.
0 or OFF: disable selected channel auto noise.
1 or ON: enable selected channel auto noise.