Calculate[1..n][:wdm]:channel: center:frequency – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch

Seite 429

Advertising
background image

SCPI-Befehlsreferenz

OSA

421

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the <Center> parameter.

The <Center> parameter corresponds to a valid

channel center frequency in hertz.

The

CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ

uency? MIN and

CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ

uency? MAX queries can be used to determine

valid channel center frequency range.

Example(s)

CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ

CALC:WDM:CHAN:SEL "ITU_22"

CALC:WDM:CHAN:CENT:FREQ? Returns

1.921750E+014

CALC:WDM:CHAN:CENT:FREQ 193.4145 THZ

CALC:WDM:CHAN:CENT:FREQ? Returns

1.934145E+014

See Also

:CALCulate[1..n][:WDM]:CHANnel[:DEFine]

:CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ

uency?

:CALCulate[1..n][:WDM]:CHANnel:CENTer[:WAV

Elength]

:CALCulate[1..n][:WDM]:CHANnel:SELect

:CALCulate[1..n][:WDM]:CHANnel:

CENTer:FREQuency

Advertising