EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 389

SCPI-Befehlsreferenz
OSA
381
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:DISTance:[WAVelength]
Description
This command sets the wavelength distance
from peak to center of noise region for noise
measurement of the WDM analysis default
channel.
At *RST, default channel noise measurement
distance is set to 100.0 GHz.
Syntax
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:[WAVelength]<wsp><Distance[<ws
p>M]>|MAXimum|MINimum|DEFault
Parameter(s)
Distance:
The program data syntax for <Distance> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is M. The
<Distance> special forms MINimum, MAXimum
and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.