Sense[1..n]:correction:offset [:magnitude – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 571
Advertising

SCPI-Befehlsreferenz
OSA
563
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Offset> parameter.
The <Offset> parameter corresponds to a valid
power offset value.
The
SENSe[1..n]:CORRection:OFFSet[:MAGNitude]?
MIN and
SENSe[1..n]:CORRection:OFFSet[:MAGNitude]?
MAX queries can be used to determine valid
power offset range.
Example(s)
SENS:CORR:OFFS:MAGN 0.5 DB
UNIT:RAT DB
SENS:CORR:OFFS:MAGN? Returns
5.000000E-001
See Also
:SENSe[1..n]:WAVelength:OFFSet
:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]?
:SENSe[1..n]:CORRection:OFFSet
[:MAGNitude]