Calculate[1..n][:wdm]:channel: nselect – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch

Seite 470

Advertising
background image

SCPI-Befehlsreferenz

462

FTB-5240S/S-P/BP

Product-Specific Commands—Description

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

The <Select> parameter corresponds to a valid

channel setup index to select. The channel index

cannot be zero.

The CALCulate[1..n][:WDM]:CHANnel:COUNt?

query can be used to determine valid index

range.

Example(s)

CALC:WDM:CHAN:DEL:ALL

CALC:WDM:CHAN:DEF "C_001",1525.000 NM

CALC:WDM:CHAN:NSEL 1

CALC:WDM:CHAN:SEL? Returns "C_001"

See Also

:CALCulate[1..n][:WDM]:CHANnel:COUNt?

:CALCulate[1..n][:WDM]:CHANnel[:DEFine]

:CALCulate[1..n][:WDM]:CHANnel:NSELect?

:CALCulate[1..n][:WDM]:CHANnel:SELect

:CALCulate[1..n][:WDM]:DATA:CHANnel:NSELec

t

:CALCulate[1..n][:WDM]:CHANnel:

NSELect