Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:AUTO
Description
This command controls activation of i-InBand
noise measurement for WDM analysis default
channel.
At *RST, auto noise is set to off (disabled).
Syntax
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO<wsp><Auto>
Parameter(s)
Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of auto noise measurement.
0 or OFF: disable default channel auto noise.
1 or ON: enable default channel auto noise.