SCPI-Befehlsreferenz

OSA

515

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:DATA:

SIGnalPower:FLATness?

Description

This query returns computed WDM analysis

global result for signal power flatness.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:DATA:SIGnalPower:FLA

Tness?

Parameter(s)

None

Response Syntax

<Flatness>

Response(s)

Flatness:
The response data syntax for <Flatness> is

defined as a <NR3 NUMERIC RESPONSE DATA>

element.

The <Flatness> response corresponds to the

computed signal power flatness.

Example(s)

<Do measurement>

UNIT:RAT DB

CALC:DATA:SIGP:FLAT? Returns 3.118000E+001

Notes

Special NAN (not a number) value

-2251799813685248 is returned if analysis result

could not be computed.

See Also

:CALCulate[1..n][:WDM]:DATA:OSNR:FLATness?

:CALCulate[1..n][:WDM]:DATA:OSNR:MEAN?

:CALCulate[1..n][:WDM]:DATA:SIGnalPower:ME

AN?

:CALCulate[1..n][:WDM]:DATA:TPOWer?