SCPI-Befehlsreferenz

566

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:SENSe[1..n]:FREQuency:STARt

Description

This command sets instrument sweep start

frequency.

At *RST, this value is set to 190.9506 THz.

Syntax

:SENSe[1..n]:FREQuency:STARt<wsp><Start[<

wsp>HZ]>|MAXimum|MINimum|DEFault

Parameter(s)

Start:
The program data syntax for <Start> is defined

as a <numeric_value> element followed by an

optional <SUFFIX PROGRAM DATA> element.

The allowed <SUFFIX PROGRAM DATA>

element is HZ. The <Start> special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the <Start> parameter.

The <Start> parameter corresponds to a valid

sweep start frequency value.

The SENSe[1..n]:FREQuency:STARt? MIN and

SENSe[1..n]:FREQuency:STARt? MAX queries

can be used to determine valid sweep start

frequency range.