Trace:bandwidth|bwidth:resolution – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 598

SCPI-Befehlsreferenz
590
FTB-5240S/S-P/BP
Product-Specific Commands—Description
:TRACe:BANDwidth|BWIDth:RESolution?
Description
This query returns the resolution bandwidth of
the wavelength range for the specified trace.
This query has no associated *RST condition.
Syntax
:TRACe:BANDwidth|BWIDth:RESolution?<wsp>
<TraceName>
Parameter(s)
TraceName:
The program data syntax for <TraceName> is
defined as a <STRING PROGRAM DATA>
element.
The <TraceName> parameter corresponds to
the name of the trace to select.
Response Syntax
<RBW>
Response(s)
RBW:
The response data syntax for <RBW> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
The <RBW> response corresponds to the
resolution bandwidth of the current wavelength
range of the trace expressed in meter.
Example(s)
<Do measurement>
TRAC:BAND:RES? "TRC1" Returns 6.2015E-011
Notes
The only valid trace name is "TRC1".
Trace data is available only if trace analysis was
performed.
See Also
:CALCulate[1..n][:WDM]:DATA:CHANnel:ENBW?