background image

SCPI-Befehlsreferenz

420

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

CENTer:FREQuency

Description

This command sets the nominal center

frequency of the selected WDM analysis channel.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ

uency<wsp><Center[<wsp>HZ]>|MAXimum

|MINimum|DEFault

Parameter(s)

Center:
The program data syntax for <Center> is

defined as a <numeric_value> element

followed by an optional <SUFFIX PROGRAM

DATA> element. The allowed <SUFFIX

PROGRAM DATA> element is HZ. The <Center>

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

Advertising