EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) Benutzerhandbuch
Seite 551

SCPI-Befehlsreferenz
OSA
543
Product-Specific Commands—Description
:MEMory:TABLe:DATA?
Description
This query returns the channel results in a
"row-column" format for the specified table. The
list of column is specified using the
:MEMory:TABLe:DEFine command. The number
of rows is available using :MEMory:TABLe:POINt?
command.
This query has no associated *RST condition.
Syntax
:MEMory:TABLe:DATA?<wsp><TableName>
Parameter(s)
TableName:
The program data syntax for <TableName> is
defined as a <STRING PROGRAM DATA>
element.
The <TableName> parameter corresponds to
the name of the table to select.
Response Syntax
<Table>
Response(s)
Table:
The response data syntax for <Table> is defined
as a <DEFINITE LENGTH ARBITRARY BLOCK
RESPONSE DATA> element.
The <Table> response contains an array of
channel results. Each string line corresponds to a
row in the table. Each row is composed of
column where the entries are specified in the
:MEMory:TABLe:DEFine command. The column
order is preserved. Unrecognized column
definition will produce empty result.