temp-chart Software

temp-gard Hardware

10

Barriers and Heat Sinks

Made from stainless steel the Thermal Barrier provides the thermal and mechanical
protection necessary for the data logger to travel through the hostile environment of an
industrial oven.

Ceramic insulation covered by a ceramic fiber cloth provides the primary thermal
protection. Heat sinks provide additional, secondary protection allowing the system to
operate at high temperature for extended periods. The heat sinks are filled with a non-
toxic, non-flammable phase changing material. This material absorbs the heat and
maintains a temperature of approx. +50 °C until all the material has changed from solid to
liquid state.

Note:

The ceramic insulation material will absorb moisture if stored in a damp

environment. Although this will not result in damage to the thermal barrier, its performance
will be degraded until the moisture is removed.

A steel pad on the barrier’s lid provides a convenient stage area for magnetically attached
probes.

Thermocouple Probes

Thermocouple probes utilize the Seebeck effect which, discovered in the 19

th

century,

results in a voltage proportional to temperature being produced at the junction of any two
dissimilar metals.

The actual voltage measured is proportional to the temperature difference between the
thermocouples “Hot” and “Cold” junctions, the “Hot” junction being the measurement
junction, and the “Cold” junction being the junction of thermocouple and measurement
instrumentation.

The practical implementation of thermocouples requires sophisticated electronics to
eliminate potential measurement errors. These potential errors include poor linearity over
the measurement range, and inaccuracy due to temperature variations at the Cold
reference junction.

To accommodate these the electronics in the measuring system must simulate a
temperature of zero degrees at this second junction, as well as compensating for any
non linearity over the range of the thermocouple operation.