17

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

DANGER!

The ProcessCalibrator must be switched off before connecting or

disconnecting any electrical connection. In addition, the supply voltage

marked on the power supply unit must match the local mains voltage.

The measuring inputs must not be electrically overloaded (see chapter

3 "Specifications") and if the test item has its own power supply, the

internal 24 V supply must be switched off via the menu.

WARNING!

If the ProcessCalibrator is set for reading from test items with voltage

outputs (e.g. 0 ... 1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V) and no test item is

connected to the measuring input (voltage), then a non-zero value will

be shown for the test item on the display. This is not an error, it is simply

due to the electrical design of the measuring input.

The internal 24 V voltage supply must not be short-circuited nor should

the max. output current through the snubber circuit exceed 50 mA. (In

addition, it should not drop below 20 mA, in order to ensure an accurate

current measurement.)

4.3.1 Electrical connection of potential-free pressure switches

WARNING!

Only potential-free (passive) switches, as shown in the drawing

and using the supplied test cable, should be connected to the

ProcessCalibrator.

A current or voltage input could damage the CPH6000.

4. Design and function

Normally open /

closed

Low-/High-window

Type of switch

Potential-free

contact