WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
d) Functions: Operating functions such as
■
Tare: Offset correction of the reference pressure value
■
Min/Max: Minimum/Maximum memory
■
Alarm: Min/Max alarm (visual and audible)
■
Filter: Damping/smoothing of the reference sensor signal
For further information see chapter 7.4.1 "SETUP additional menu items: Functions"
e) CPH-Info: General CPH6000 instrument data such as
■
Calibration dates for the electrical measurement inputs
■
Firmware number
■
Serial No. of the device
For further information see chapter 7.4.2 "SETUP additional menu items: CPH-INFO"
f) ReferenceSensor: Data for the currently-connected reference pressure sensor such
as
■
Measuring range
■
Accuracy class
■
Sensor measurement type
■
Information in the event of reference sensor overpressure
■
Calibration dates for the reference sensor
For further information see chapter 7.4.3 "SETUP additional menu item: Reference-
Sensor"
g) Ref. Sensor-list:
■
List of the stored reference sensors that can be attached and are calibrated.
For further information see chapter 7.4.4 "SETUP additional menu item: Ref. Sensor-
list"
h) CPH-Configuration:
■
Info: on battery level
■
Setting options from: Menu language, system time / system clock, display brightness,
powersave function (automatic energy saving mode; see chapter 7.4.5 "SETUP addition-
al menu item: CPH-Configuration")
■
Input options:
- the ambient temperature during the calibration
- the height difference between reference pressure sensor and test item
(see chapter 6.1 "Requirements for test assemblies with the CPH6000").
For further information see chapter 7.4.5 "SETUP additional menu item: CPH-Config-
uration"
i)
Interface
■
Switching between USB and RS-232 interfaces incl. entry of baud rate
For further information see chapter 7.4.6 "SETUP additional menu item: Interface"
j)
CLEAR CalData:
■
Delete all stored calibration data (clear and reset all memory locations)
For further information see chapter 7.4.7 "SETUP additional menu items: CLEAR
CalData"
k) CPH switch off:
■
Switching off the CPH6000 ProcessCalibrator
For further information see chapter 7.2.2 "Switching off the model CPH6000
ProcessCalibrator"
7. Commissioning, operation