WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

d) Functions: Operating functions such as

■

Tare: Offset correction of the reference pressure value

■

Min/Max: Minimum/Maximum memory

■

Alarm: Min/Max alarm (visual and audible)

■

Filter: Damping/smoothing of the reference sensor signal

For further information see chapter 7.4.1 "SETUP additional menu items: Functions"

e) CPH-Info: General CPH6000 instrument data such as

■

Calibration dates for the electrical measurement inputs

■

Firmware number

■

Serial No. of the device

For further information see chapter 7.4.2 "SETUP additional menu items: CPH-INFO"

f) ReferenceSensor: Data for the currently-connected reference pressure sensor such

as

■

Measuring range

■

Accuracy class

■

Sensor measurement type

■

Information in the event of reference sensor overpressure

■

Calibration dates for the reference sensor

For further information see chapter 7.4.3 "SETUP additional menu item: Reference-

Sensor"

g) Ref. Sensor-list:

■

List of the stored reference sensors that can be attached and are calibrated.

For further information see chapter 7.4.4 "SETUP additional menu item: Ref. Sensor-

list"

h) CPH-Configuration:

■

Info: on battery level

■

Setting options from: Menu language, system time / system clock, display brightness,

powersave function (automatic energy saving mode; see chapter 7.4.5 "SETUP addition-

al menu item: CPH-Configuration")

■

Input options:

- the ambient temperature during the calibration

- the height difference between reference pressure sensor and test item

(see chapter 6.1 "Requirements for test assemblies with the CPH6000").

For further information see chapter 7.4.5 "SETUP additional menu item: CPH-Config-

uration"

i)

Interface

■

Switching between USB and RS-232 interfaces incl. entry of baud rate

For further information see chapter 7.4.6 "SETUP additional menu item: Interface"

j)

CLEAR CalData:

■

Delete all stored calibration data (clear and reset all memory locations)

For further information see chapter 7.4.7 "SETUP additional menu items: CLEAR

CalData"

k) CPH switch off:

■

Switching off the CPH6000 ProcessCalibrator

For further information see chapter 7.2.2 "Switching off the model CPH6000

ProcessCalibrator"

7. Commissioning, operation