WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

The digital instrument and the reference pressure sensor are connected to each other

electrically using a separate connecting cable. When exchanging the sensor, use the

8-pin plug connector at the sensor.

In order to make the electrical connection for a CPT6000 reference pressure sensor,

bring together the corresponding cable plug connector to the sensor (in line with the

orientation guide) and secure it using the connection sleeve (screw in the connection

sleeve clockwise without too much force). To loosen the connection, the connection

sleeve must be turned anti-clockwise. To disconnect the sensor, do not pull on the cable,

but rather only on the connector body.

For the connection to the CPH6000, the other end of the cable must also be brought

together in line with the orientation guide and secured using the connection sleeve

(screw in the connection sleeve without too much force). To loosen the connection, the

connection sleeve must be turned anti-clockwise. To disconnect the CPH6000, do not

pull on the cable, but rather only on the connector body.

2. Electrical connection using the extension cable for remote operation/use of the

CPT6000 pressure sensors

CAUTION!

Only ever use the original WIKA extension cable for the remote

operation of CPT6000 reference pressure sensors, and never more

than a single cable.

4. Design and function

The connector must be disconnected

and connected, respectively, in accord-

ance with the instructions in 1.

4.4.3 Mechanical connection of CPT6000 reference pressure sensors to the

CPH6000

To make the mechanical connection of the CPT6000 reference pressure sensor, it must

be placed, connecting thread first, in the hexagonal sensor bracket of the instrument,

so that the hexagonal anti-rotation stop of the CPT6000 sits accordingly in the sensor

bracket. Subsequently, the sensor can be secured with the hand-operated, quick-

connect mechanism. (tighten = turn clockwise; release = turn anti-clockwise)