Commissioning, operation – WIKA CPH6000 Benutzerhandbuch

Seite 38

Advertising
background image

38

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

Functions
CPH-Info
ReferenceSensor
Ref. Sensor-list
CPH-Configuration
Interface
CLEAR CalData

CPH switch off

a

b

h

g

f

e

d

c

i

j

k

a) MEASURING: Operating mode MEASURING

To measure working/process pressures

For comparative measurements and/or calibrations (without data recording) of mechani-

cal* and electrical pressure measuring instruments (supply and display of the test item

through the CPH6000)

For further information see chapters 7.3.1 "MEASURING mode" and 7.3.2 "MEASUR-

ING mode (with test item)"

b) CALIBRATION: Operating mode CALIBRATION

For on-site calibration of mechanical* and electrical pressure measuring instruments

(without PC). In this case the data sets (for up to 16 test items, each with up to 32 test

points including date and time) are recorded within the CPH6000.

For further information see chapter 7.3.3 "CALIBRATION mode"

* For mechanical indicating instruments, the test item‘s measured value must be entered via the numeric keypad.

The calibration data can be transferred into printable certificates using

either PrintCal or EasyCal software.

c) SWITCH-TEST: Operating mode SWITCH-TEST

For the easy checking of pressure switches, including automatic calculation of the switch

hysteresis.

For further information see chapter 7.3.7 "SWITCH-TEST mode"

7. Commissioning, operation

7.2.3 Contents of the SETUP MENUS