Commissioning, operation, Shor t inf o – WIKA CPH6000 Benutzerhandbuch

Seite 41

Advertising
background image

41

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

Customer-specific unit; with respect to bar

(input via numeric keypad)

Display resolution in operating mode

(back with

)

Select and confirm (standard units) via

Shor

t inf

o:

XXX

Current cursor position; alter via

◁▷ ◀▶

Parameter selection from list or menu via

0.00

Parameter input via numeric keypad

ENTER

Input confirmation

CLEAR

Clear entry

In order to switch the instrument into "MEASURING" mode, follow the instructions on

the previous page.

The following is a more detailed explanation of point "2. Preparation for

MEASURING"

a) Test item type and test item measurement signal: [----] for measurement without test

item

b) R.-Start/Range-End:

Start of measuring range and end of measuring range of the currently to calibrate test

item

c) Unit and resolution (submenu)

7. Commissioning, operation

d) Measurement uncertainty of the test item in % FS (i.e. of the span) or % rd (i.e. of the

measured value)

e) Measurement type for the test item (relative or absolute)
f) Test medium (pneumatic -> gas or hydraulic -> oil)
g) Voltage supply for test item (on/off) [If no external supply is required for the test

item, "OFF" should be selected to conserve energy]

Advertising