Commissioning, operation – WIKA CPH6000 Benutzerhandbuch

Seite 53

Advertising
background image

53

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

ENTER

..Functions

Offset pressure

Ref.:

Tare:

Min:
Max:

Alarm:

Filter:

Ref. = rel. Pressure


0.000

0.000
0.000

ON

10.00

-1.00

0

0.000

bar

bar

bar
bar

bar
bar

a

c

b

d

e

7.4 SETUP additional menu items
7.4.1 SETUP additional menu item: Functions

7. Commissioning, operation

a) Ref.:

Current measured value of the connected CPT6000 reference pressure sensor

b) Tare:

Offset function that adjusts the current measured value. The value entered is

added to the current measured value.

(e.g. Ref. 0.000 and Tare: 1.000 --> [new] Ref. 1.000)

c) Min/Max:

Minimum and maximum value memory

The memory is reset by highlighting the value with the cursor (via

) and then

pressing the CLEAR button.

d) Alarm:

Audible and visible alarm function.

upper alarm limit: ≥ bar

lower alarm limit: ≤ bar

If the current measured value goes outside the set alarm limits, an intermittent

alarm tone sounds and the lower display line blinks.

Activation via:

Move the cursor to the field next to the word Alarm that reads <OFF> and via

toggle it to <ON>.

Deactivation via:

toggle back to <OFF>

Select menu item

Input via numerical keypad

Confirmation of the input

(Clear deletes the input, or

resets the MIN/MAX memory)