WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
ENTER
..Functions
Offset pressure
Ref.:
Tare:
Min:
Max:
Alarm:
Filter:
Ref. = rel. Pressure
≥
≤
0.000
0.000
0.000
ON
10.00
-1.00
0
0.000
bar
bar
bar
bar
bar
bar
a
c
b
d
e
7.4 SETUP additional menu items
7.4.1 SETUP additional menu item: Functions
7. Commissioning, operation
a) Ref.:
Current measured value of the connected CPT6000 reference pressure sensor
b) Tare:
Offset function that adjusts the current measured value. The value entered is
added to the current measured value.
(e.g. Ref. 0.000 and Tare: 1.000 --> [new] Ref. 1.000)
c) Min/Max:
Minimum and maximum value memory
The memory is reset by highlighting the value with the cursor (via
) and then
pressing the CLEAR button.
d) Alarm:
Audible and visible alarm function.
upper alarm limit: ≥ bar
lower alarm limit: ≤ bar
If the current measured value goes outside the set alarm limits, an intermittent
alarm tone sounds and the lower display line blinks.
Activation via:
Move the cursor to the field next to the word Alarm that reads <OFF> and via
toggle it to <ON>.
Deactivation via:
toggle back to <OFF>
Select menu item
Input via numerical keypad
Confirmation of the input
(Clear deletes the input, or
resets the MIN/MAX memory)