Commissioning, operation – WIKA CPH6000 Benutzerhandbuch

Seite 57

Advertising
background image

57

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

ENTER

a

b

c

d

e

f

g

h

7.4.5 SETUP additional menu items: CPH-Configuration

7. Commissioning, operation

This menu section lists general instrument settings, such as:

a) Input possibility for an ambient temperature

b) Input possibility for a height difference between reference pressure sensor and test

item, used in the automatic correction (deduction of a head pressure)

c) Selection of the menu language (German/English/French/Spanish/Italian)

d) Date of the system clock (Year/Month/Day)

e) Time of the system clock (Hours/Minutes/Seconds)

f) Brightness of the backlighting of the display

g) Powersave function (automatic switch-off delay for backlight and internal 24 V test

item power supply).

If the instrument is idle for the set switch-off delay period (no keys pressed and

no digital communication), then the backlight and the 24 V test item power supply

will be switched off, until any key is pressed (except ON/OFF) or the instrument is

addressed over the digital interface.

h) Current battery level

At 10 % the low battery charge warning, "low BAT", is shown in the display.

Select menu item

Input via numerical keypad

Confirmation of the input

(Clear deletes the input)